About the SciCan STATIM 2000
Description:
The SciCan STATIM 2000 cassette autoclave is a compact, countertop steam-sterilizer engineered for rapid instrument turnaround in dental, medical, and body-art practices. Instead of a conventional cylindrical pressure chamber, the STATIM uses a thin aluminum cassette that slides into a 19.5 × 16.3 × 5.9-inch chassis. Just 2.6 fl oz (≈77 mL) of distilled water is drawn from an internal 1.06-gallon reservoir for each cycle, flash-heated in a copper boiler, then pulsed through the cassette under microprocessor control. The low thermal mass of the cassette plus SciCan’s Dri-Tec™ vacuum exhaust lets an unwrapped load reach 135 °C in 15 seconds and complete sterilization in 3.5 minutes, with a total cycle of six minutes including drying; wrapped and rubber/plastic cycles are likewise pre-programmed and selectable from the soft-touch membrane panel. Integrated sensors monitor chamber temperature, pressure, and water level, while the cassette seal, bacterial retentive air filters, and printer/RS-232 port are all user-serviceable. A spring-loaded pressure-relief valve set at 43.5 psi provides fail-safe protection, and up to 30 cycles’ worth of data can be output for compliance records. The whole unit weighs only 46 lb dry and runs from a standard 110 V, 11 A outlet (220-240 V, 6 A models are also available).
Economic Advantage:
When it debuted, the STATIM 2000 dramatically cut the “idle inventory” of critical handpieces and scalers: an office could sterile-reprocess a small tray in roughly one-third the time of even a fast, benchtop chamber autoclave, freeing staff from instrument bottlenecks or the cost of keeping extra sets on hand. Its cassette architecture also consumes less than a cup of water per run and vents only cooled steam, so it needs neither a plumbed drain nor an HVAC hood—making it inexpensive to install beside the operatory and inexpensive to operate over its life cycle. The combination of six-minute turnaround, low utility use, and verified sterilization prints gave the STATIM series a total-cost advantage that competitors could not match for more than a decade.
Price History
Specifications:
External dimensions (L × W × H) | |
---|---|
Cassette internal size | 11 in × 7.1 in × 1.4 in |
Sterilization chamber volume | 0.48 gal (1.8 L) |
Reservoir capacity | 1.06 gal (4 L) |
Dry weight | 46 lb |
Electrical rating | 110 V AC, 60 Hz, 11 A (alt. 220-240 V, 50/60 Hz, 6 A) |
Unwrapped cycle | 135 °C for 3.5 min; total 6 min |
Wrapped cycle | 135 °C for 10 min; total 14 min |
Rubber/plastic cycle | 121 °C for 15 min; total 20 min |
Pressure-relief valve setting | 43.5 psi |
Minimum reservoir charge | 18.6 fl oz (550 mL) |
Ambient operating range | 41 °F – 104 °F (5 °C – 40 °C) |
Sound level | 56 dB (A) average, 65 dB peak |